|a Instrumentacijsko EKG pojačalo s istosmjerno vezanim ulaznim stupnjem :
|b završni rad /
|c Maja Barić ; [mentor Ratko Magjarević].
|
246 |
1 |
|
|a DC coupled ECG instrumentation amplifier
|i Naslov na engleskom:
|
260 |
|
|
|a Zagreb,
|b M. Barić,
|c 2015.
|
300 |
|
|
|a 38 str. ;
|c 30 cm +
|e CD-ROM
|
502 |
|
|
|b preddiplomski studij
|c Fakultet elektrotehnike i računarstva u Zagrebu
|g smjer: Elektroničko i računalno inženjerstvo, šifra smjera: 35, datum predaje: 2015-06-12, datum završetka: 2015-07-13
|
520 |
3 |
|
|a Sažetak na hrvatskom: U ovom radu rješavao se problem projektiranja jednokanalnog EKG pojačala. Projektirao se prema ulaznim zahtjevima prema kojima je predviđeno napajanje pojačala baterijom nominalnog napona 3,7 V, mora imati 2 načina rada: dijagnostički (0,05 – 120 Hz) i monitoriranje (1 – 45 Hz), a ulazni filtar mora biti niskopropusni. Kao ulazni filtar odabran je niskopropusni RC filtar. Nakon pojačanja u instrumentacijskom pojačalu, guše se niske frekvencije u Besselovom visokopropusnom filtru. Na kraju se signal dodatno pojačava u neinvertirajućem pojačalu.
Dobiveno je pojačalo karakteristika bliskih zahtijevanim. Izmjeren je EKG na ispitaniku pri čemu je dobiven očekivani signal na izlazu.
|
520 |
3 |
|
|a Sažetak na engleskom: This thesis is based on a problem of designing a single – channel electrocardiogram. Its design is based on given specifications which include 3,7 V single power supply, low – pass filter in its input stage and ability to work in two different modes chosen by the user: diagnosting mode (0,05 – 120 Hz) and monitoring mode (1 – 45 Hz). The chosen input stage filter is low – pass RC filter. After amplifying the signal in an instrumentation amplifier, low frequencies are attenuated in a high – pass Bessel filter. Finally, the signal is amplified in a non – inverting amplifier.
As a result, the designed ECG amplifier was confirmed to have very close specifications to the ones requested. Finally, ECG was measured on a test subject which resulted in expected output signals.
|
653 |
|
1 |
|a EKG
|a instrumentacijsko pojačalo
|a visokopropusni Besselov filtar
|a niskopropusni RC filtar
|a smetnje
|
653 |
|
1 |
|a ECG
|a instrumentation amplifier
|a high – pass Bessel filter
|a low – pass RC filter
|a noise
|