Description of interference in the measurement of troponin T by a high-sensitivity method

Measurement of high-sensitivity troponin T (hs-TnT) has become an essential step in the diagnosis of acute myocardial infarction.This high-sensitivity method allows quantifying the concentration of troponin T in blood of healthy subjects with a lower inaccuracy compared toprevious reagent generation...

Full description

Permalink: http://skupni.nsk.hr/Record/nsk.NSK01001069016/Details
Matična publikacija: Biochemia medica (Online)
29 (2019), 2 ; str. 413-419
Glavni autori: Aliste-Fernández, Miguel (Author), Sole-Enrech, Gemma, Cano-Corres, Ruth, Teodoro-Marin, Silvia, Berlanga-Escalera, Eugenio
Vrsta građe: e-članak
Jezik: eng
Predmet:
Online pristup: https://doi.org/10.11613/BM.2019.021001
Biochemia medica (Online)
Hrčak
LEADER 03169naa a22004094i 4500
001 NSK01001069016
003 HR-ZaNSK
005 20200907142220.0
006 m d
007 cr||||||||||||
008 200710s2019 ci d |o |0|| ||eng
024 7 |2 doi  |a 10.11613/BM.2019.021001 
035 |a (HR-ZaNSK)001069016 
040 |a HR-ZaNSK  |b hrv  |c HR-ZaNSK  |e ppiak 
041 0 |a eng  |b eng 
042 |a croatica 
044 |a ci  |c hr 
080 1 |a 577  |2 2011 
080 1 |a 616  |2 2011 
100 1 |a Aliste-Fernández, Miguel  |4 aut  |9 HR-ZaNSK 
245 1 0 |a Description of interference in the measurement of troponin T by a high-sensitivity method  |h [Elektronička građa] /  |c Miguel Aliste-Fernández, Gemma Sole-Enrech, Ruth Cano-Corres, Silvia Teodoro-Marin, Eugenio Berlanga-Escalera. 
300 |b Graf. prikazi. 
504 |a Bibliografija: 16 jed. 
504 |a Summary. 
520 |a Measurement of high-sensitivity troponin T (hs-TnT) has become an essential step in the diagnosis of acute myocardial infarction.This high-sensitivity method allows quantifying the concentration of troponin T in blood of healthy subjects with a lower inaccuracy compared toprevious reagent generations. However, the presence of certain compounds in the sample may interfere with the result. We present a patient whohad repeatedly high concentrations of hs-TnT in the serum sample that did not agreed with the signs and symptoms. In addition, ultrasensitive troponinI concentration was undetectable.Materials and methods: To investigate the presence of an interfering compound, different analysis were carried out. In order to discard macrocomplexes in the sample, the serum was precipitated with polyethylene glycol. In addition, the serum was incubated with Scantibodies HeterophilicBlocking Tube, which can block heterophilic antibodies. Finally, a size exclusion chromatography of the sample was performed by the manufacturer.What happened: The interfering substance was allocated into fractions with proteins of 150kDa, corresponding to high molecular weight proteinslike immunoglobulin G (IgG). This compound was responsible for the falsely elevated hs-TnT results and it affected only the high-sensitivity methods.Main lesson: The detected interfering compound was probably an IgG. This type of interference must be kept in mind in front of discordant results,even if they are extremely rare. Therefore, interdisciplinary cooperation between clinicians, laboratory and manufacturer is essential. 
653 0 |a Troponin T  |a Mjerenje  |a Smetnje  |a Infarkt miokarda 
700 1 |a Sole-Enrech, Gemma  |4 aut  |9 HR-ZaNSK 
700 1 |a Cano-Corres, Ruth  |4 aut  |9 HR-ZaNSK 
700 1 |a Teodoro-Marin, Silvia  |4 aut  |9 HR-ZaNSK 
700 1 |a Berlanga-Escalera, Eugenio  |4 aut  |9 HR-ZaNSK 
773 0 |t Biochemia medica (Online)  |x 1846-7482  |g 29 (2019), 2 ; str. 413-419  |w nsk.(HR-ZaNSK)000655372 
981 |b Be2019  |b B04/19 
998 |b dalo2009 
856 4 0 |u https://doi.org/10.11613/BM.2019.021001 
856 4 0 |u https://www.biochemia-medica.com/en/journal/29/2/10.11613/BM.2019.021001  |y Biochemia medica (Online) 
856 4 1 |y Digitalna.nsk.hr 
856 4 0 |u https://hrcak.srce.hr/221086  |y Hrčak