Contribution of the laboratory to a diagnosis process by sequential reflective testing: Paraprotein interference on a direct bilirubin assay

Errors in laboratory medicine occur in the preanalytical, analytical, and postanalytical phases. The errors are mostly detected in the preanalytical period. However, analytical errors are still an important source of error, despite their frequency is reduced significantly in years thanks to developm...

Full description

Permalink: http://skupni.nsk.hr/Record/nsk.NSK01001163320/Details
Matična publikacija: Biochemia medica (Online)
31 (2021), 2 ; str. 351-358
Glavni autori: Yilmaz, Niyazi Samet (Author), Sen, Bayram, Gulbahar, Ozlem
Vrsta građe: e-članak
Jezik: eng
Online pristup: https://doi.org/10.11613/BM.2021.020801
Elektronička verzija članka
Elektronička verzija članka
Elektronička verzija članka
LEADER 03192naa a22003614i 4500
001 NSK01001163320
003 HR-ZaNSK
005 20230213141809.0
006 m d
007 cr||||||||||||
008 230213s2021 ci |o |0|| ||eng
024 7 |2 doi  |a 10.11613/BM.2021.020801 
035 |a (HR-ZaNSK)001163320 
040 |a HR-ZaNSK  |b hrv  |c HR-ZaNSK  |e ppiak 
041 0 |a eng 
042 |a croatica 
044 |a ci  |c hr 
080 1 |2 2011 
100 1 |a Yilmaz, Niyazi Samet  |4 aut  |9 HR-ZaNSK 
245 1 0 |a Contribution of the laboratory to a diagnosis process by sequential reflective testing: Paraprotein interference on a direct bilirubin assay  |h [Elektronička građa]  |c Niyazi Samet Yilmaz, Bayram Sen, Ozlem Gulbahar. 
300 |b Ilustr. 
504 |a Bibliografija: 
504 |a Summary. 
520 |a Errors in laboratory medicine occur in the preanalytical, analytical, and postanalytical phases. The errors are mostly detected in the preanalytical period. However, analytical errors are still an important source of error, despite their frequency is reduced significantly in years thanks to developments in laboratories. In this case, an analytical error was noticed during the verification of a patient"s results. The direct bilirubin of a 66-year-old male patient admitted to the emergency department was higher than the total bilirubin. The patient"s symptoms were fatigue and dyspnoea. Albumin and haemoglobin (Hb) concentrations of the patient were significantly low. After considering the patient"s demographics and laboratory results, the laboratory specialist suspected a paraproteinemia interference. Total protein was performed as a reflective test. The albumin/globulin ratio was reversed. Thereafter, serum protein electrophoresis (SPEP) and immunofixation electrophoresis (IFE) were performed as another reflective tests, respectively. SPEP and IFE results were in favour of monoclonal gammopathy. The patient was directed to a haematologist, underwent a bone marrow biopsy, and the result was reported as Waldenstrom"s macroglobulinemia with plasma cell differentiation expressing IgM-Kappa. The patient went on a chemotherapy protocol, and his condition has been improved in subsequent months. Detection of analytical errors is of great importance, like in our case, and may be used as a tool to identify patients who have not yet been diagnosed. The laboratory specialist must dominate the entire process of each test in the laboratory, be aware of the limitations of tests, and turn these disadvantages into advantages when necessary. 
700 1 |a Sen, Bayram  |4 aut  |9 HR-ZaNSK 
700 1 |a Gulbahar, Ozlem  |4 aut  |9 HR-ZaNSK 
773 0 |t Biochemia medica (Online)  |x 1846-7482  |g 31 (2021), 2 ; str. 351-358  |w nsk.(HR-ZaNSK)000655372 
981 |b Be2021 
856 4 0 |u https://doi.org/10.11613/BM.2021.020801 
856 4 0 |u https://www.biochemia-medica.com/en/journal/31/2/10.11613/BM.2021.020801  |y Elektronička verzija članka 
856 4 0 |u https://www.biochemia-medica.com/en/journal/31/2/10.11613/BM.2021.020801/fullArticle  |y Elektronička verzija članka 
856 4 1 |y Digitalna.nsk.hr 
856 4 0 |u https://hrcak.srce.hr/258900  |y Elektronička verzija članka